Trig Holdoff - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

7  Trigger Functions
Trigger
Example
Notes
Preset
State Saved
Min
Max
Default Unit
Initial S/W Revision

Trig Holdoff

Sets the holdoff time between triggers. When the trigger condition is satisfied, the trigger occurs, the delay
begins, and the holdoff time begins. New trigger conditions will be ignored until the holdoff time expires.
For a free-running trigger, the holdoff value is the minimum time between triggers.
Key Path
Remote Command
Example
Dependencies
Preset
State Saved
Min
Max
Default Unit
Initial S/W Revision
376
:TRIGger[:SEQuence]:ATRigger:STATe?
TRIG:ATR:STAT ON
TRIG:ATR 100 ms
The "time that the analyzer will wait" starts when the analyzer is ready for a trigger, which may be
hundreds of ms after the data acquisition for a sweep is done. The "time" ends when the trigger
condition is satisfied, not when the delay ends.
Off, 100 ms
Saved in instrument state
1 ms
100 s
s
Prior to A.02.00
Trigger, Auto/Holdoff
:TRIGger[:SEQuence]:HOLDoff <time>
:TRIGger[:SEQuence]:HOLDoff?
:TRIGger[:SEQuence]:HOLDoff:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:HOLDoff:STATe?
TRIG:HOLD:STAT ON
TRIG:HOLD 100 ms
Unavailable if the selected Input is BBIQ.  If this is the case, the key is grayed out if it is pressed the
informational message "Feature not supported for this Input" is displayed.  If the SCPI command is
sent, the error "Settings conflict; Feature not supported for this Input" is generated.
Off, 100 ms
Saved in instrument state
0 s
0.5 s
s
Prior to A.02.00
EMI Receiver Mode Reference