Scroll Left
Scroll signal list to the left.
Key Path
Notes
Dependencies
Initial S/W Revision
Scroll Right
Scroll signal list to the right.
Key Path
Notes
Dependencies
Initial S/W Revision
Signal à Meters (Move Meters to Current Signal)
Replace Frequency (Meters) with the frequency of the selected signal.
Key Path
Remote Command
Example
Notes
Dependencies
Initial S/W Revision
Mark Signals
This menu enables you to mark signals for further processing. A ticked checkbox appears next to the signal
number to distinguish them from unmarked signals.
Key Path
Initial S/W Revision
EMI Receiver Mode Reference
Meas Setup, Signal List, Navigate
Front panel only.
This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict" warning.
A.07.00
Meas Setup, Signal List, Navigate
Front panel only.
This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict" warning.
A.07.00
Meas Setup, Signal List, Navigate
:CALCulate:FSCan:SLISt:SET:METer <integer>
CALC:FSC:SLIS:SET:MET 12 Set Freq (Meters) to frequency of Signal #12
If the signal list contains no data, an error is generated and this function cannot be performed.
If the selected signal is not in the list, an error is generated and this function cannot be performed.
This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict" warning.
A.07.00
Meas Setup, Signal List
A.07.00
8 Frequency Scan Measurement
Meas Setup
441