Delta

Sets the control mode for the selected marker to Delta and turns on the active function for setting its delta
value. If the selected marker is off, the marker is placed at the beginning of the trace.
In Delta mode, the marker result shows the relative result between the selected (Delta) marker and its
reference marker. A delta marker can be moved to any point on the X Axis by specifying its X Axis offset
from a reference marker. Its absolute Y Axis value is then the value of the trace point at that X Axis value.
Key Path
Example
Dependencies
State Saved
Initial S/W Revision
Off
Turns off the selected marker, removes the marker annunciation from the display, and turns off any active
function. Turning the marker off does not affect which marker is selected.
Key Path
Example
State Saved
Initialt S/W Revision
Properties
Accesses the marker properties menu.
Key Path
Initial S/W Revision
Select Marker
Accesses a menu that allows you to select one of 12 markers for control and function.
Key Path
Initial S/W Revision
Relative To
This sets the reference marker that the selected marker will be relative to.
EMI Receiver Mode Reference
Marker
CALC:APD:MARK:MODE DELT  Sets marker 1 to Delta.
The marker addressed by this command becomes the selected marker on the front panel.
Saved in instrument state.
A.13.00
Marker
CALC:APD:MARK:MODE OFF  Sets Marker 1 to Off.
Saved in instrument state.
A.13.00
Marker
A.13.00
Marker
A.13.00
11  APD Measurement