13 Set The Following Parameters; The Appropriate Wavelength For Your Duts; Laser Power: +3 Dbm Optical Input: Input; User Calibration: None - Keysight N4372E User Manual

Lightwave component analyzer
Table of Contents

Advertisement

3
Performance Quick Check
44
12 Select the OO measurement tab.

13 Set the following parameters:

If you have option 102 (Dual wavelength 1310 nm, 1550 nm), select

the appropriate wavelength for your DUTs.

Laser power: -1 dBm
Optical input: Input 1

User Calibration: None

Forward RF Power: 0 dBm

14 Click
, then
15 Press
to start optical power meter application.

16 Read the optical power from the power meter application.

The readout should be -1 dBm ± 0. 5 dB
17 If you have option -301 and -021:
Connect the straight-angled optical patch cord (N4373-87907) to the
optical output of an external laser, with 1310 nm or 1550 nm
wavelength and known output power, and input of the LCA Optical
Receiver. Note that the angled connector must be attached to the
optical test set.
Keysight N4372E Lightwave Component Analyzer User's Guide
to set the output power.

Advertisement

Table of Contents
loading

Table of Contents