SCPI Command Reference
E4980A/AL commands
Parameter
FAIL (Preset value)
PASS
Equivalent key
[Meas Setup] - LIMIT TABLE - BEEP - FAIL
[Meas Setup] - LIMIT TABLE - BEEP - PASS
:COMParator:BIN:CLEar
Syntax
:COMParator:BIN:CLEar
Description
Clears all of the BIN limit value settings. (No Query)
Equivalent key
[Meas Setup] - LIMIT TABLE - BIN - CLEAR TABLE
:COMParator:BIN:COUNt:CLEar
Syntax
:COMParator:BIN:COUNt:CLEar
Description
Clears all BIN counts. (No Query)
Equivalent key
[Display Format] - BIN COUNT - COUNT - RESET COUNT
:COMParator:BIN:COUNt:DATA?
Syntax
:COMParator:BIN:COUNt:DATA?
Description
Returns the comparator BIN count results (Query Only)
Query Response
<Bin 1>,<Bin 2>,<Bin 3>,<Bin 4>,<Bin 5>,<Bin 6>,<Bin 7>,<Bin 8>,<Bin 9>,<Out
of Bin>,<Aux Bin><NL^END>
Equivalent key
No equivalent key is available on the front panel.
:COMParator:BIN:COUNt[:STATe]
Syntax
:COMParator:BIN:COUNt[:STATe] {ON|OFF|1|0}
:COMParator:BIN:COUNt[:STATe]?
Description
Enables the BIN count function.
330
Description
Sounds a beep when the test is failed
Sounds a beep when the test is passed
Keysight E4980A/AL Precision LCR Meter