Powering Off The Trace 1600/1610 - Thermo Scientific TRACE 1600 Hardware Manual

Gas chromatographs
Table of Contents

Advertisement

Powering Off the TRACE 1600/1610

Thermo Scientific
To shut down the GC
1. Cool down the GC.
Note By pressing the Maintenance button, the GC cool down is automatically
carried out.
2. If you do not plan to replace the column or perform maintenance on the GC, you do not
have to lower the injector temperature.
3. Turn off the carrier gas supply at the tank.
4. Push down the power switch (breaker), located at the back of the instrument, to the
position O (down). See
Figure 73. GC Power Off
Power Switch Off Position
AC Input Connector
5. If external modules are present, unplug the power cable from the AC Input connector of
each module, and from the wall outlet.
6. If present, power-off the autosampler by switching off the main power switch, or by
unplugging the power cable from the AC input connector, and from the wall outlet.
7. Power-off all the remaining instruments.
Figure
73.
2
Performing Routine Maintenance
Powering Off the TRACE 1600/1610
TRACE 1600/1610 Hardware Manual
73