Replacing A Ssl Broken Liner - Thermo Scientific TRACE 1600 Hardware Manual

Gas chromatographs
Table of Contents

Advertisement

Replacing a SSL Broken Liner

Thermo Scientific
14. Turn the carrier gas on.
15. Set the normal injector, detector, and GC working conditions.
To replace a broken liner
Figure 80. SSL Injector: Broken Liner Replacement
Septum Cap
Ring Nut
Septum Holder/Liner Cap with Septum
Liner Seal
Liner
Base Seal
Washer
Retaining Nut
Ferrule
Capillary Column Retaining Nut
1. Put the GC in standby condition.
2. Cool the oven and injector to room temperature.
3. Turn the carrier gas off, and wait for the carrier pressure to go to zero.
3
Performing Injectors Routine Maintenance
Maintaining a Split/Splitless Injector (SSL)
Bottom Fittings
Terminal Fitting for Capillary Column
TRACE 1600/1610 Hardware Manual
91