Upgrading Equipment
Upgrading a TRACE 1600 to a TRACE 1610
588
TRACE 1600/1610 Hardware Manual
f.
Disconnect the flat cables from the connector located on the front of the door. See
Figure
612.
Figure 612. Cable Removal
7. Install the door cover of the TRACE 1610.
a. The door cover of the TRACE 1610 includes the touch screen and the cables for its
connection.
Figure 613
door where the touch screen and ground cables must be connected.
Figure 613. Cables Connections Points
shows the connection points on the internal section of the
Flat Cables
Touch Screen Connection
Point
Ground Connection
Point