Making the Duct to Couple with a DFS, IRMS, or ICP-MS Mass Spectrometer

Thermo Scientific
ATTENTION – RISQUE D'ENDOMMAGEMENT DE L'INSTRUMENT : veillez à ce que la
colonne GC ait été conditionnée avant de la brancher sur la ligne de transfert. Le matériau
libéré par la colonne (bleeding de la colonne) pendant le conditionnement risque de
contaminer le détecteur.
5. Close the front door of the GC.
6. To tune and set the ISQ Series or TSQ Series working conditions, refer to the relevant
User Guide and Hardware Manual.
To make the duct for the transfer line
1. Remove the left/right side panel according to the high resolution mass spectrometer to
couple.
a. Open the front door of the GC.
b. Use a T20 Torxhead screwdriver to loosen the left/right side panel screw from the
interior front panel. Save the screw because it will be reused later.
c. Slide the panel towards the back of the instrument up to the stop.
d. Remove the panel by pulling it outward. Be aware that the ground wire is attached to
the panel.
2. Prepare the proper duct for the transfer line inner tube.
a. Use a T20 Torxhead screwdriver to loosen the duct plate screws from the left/right
panel. See
Figure
31.
1
Coupling to a Mass Spectrometer
TRACE 1600/1610 Hardware Manual
Installation
27