Figure 338. Purge & Trap Adapter (5)
CAUTION Do not overtighten the septum cap because you might damage the septum and
affect instrument performance.
ATTENTION Ne serrez pas excessivement le capuchon du septum, car cela pourrait
endommager le septum et affecter les performances de l'instrument.
13. Close the module flap cover.
14. Connect the purge & trap adapter to the Purge & Trap system. See
Figure 339. Purge & Trap Adapter (6)
a. Using the nut and ferrule provided, connect the heated transfer line from the Purge
& Trap on the connector IN of the purge & trap adapter.
b. Using the nut and ferrule provided, connect the gas line from the injector module to
the Purge & Trap on the connector OUT of the purge & trap adapter.
15. Reconnect the column end to the injector and verify the connection point.
Installing the Purge & Trap Adapter Kit on the SSL/SSLBKF Injector
IN
OUT
TRACE 1600/1610 Hardware Manual