Thermo Scientific TRACE 1600 Hardware Manual page 334

Gas chromatographs
Table of Contents

Advertisement

7
Detectors Advanced Maintenance
Cleaning or Replacing the NPD Jet
302
TRACE 1600/1610 Hardware Manual
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
6. Unplug the power cable from the AC input connector of the NPD Thermionic Source
Power Module and from the wall outlet.
If other external modules are present, unplug the power cable from the AC Input
connector of each module, and from the wall outlet.
7. If present, move the autosampler away from the module.
8. Unplug the signal cable from the detector module. See
Figure 268. NPD Signal Cable Removal
9. Open the module flap cover.
10. Remove the thermionic source assembly cable. See
Figure 269. NPD Thermionic Source Cable Removal
Figure
268.
Figure
269.
Thermionic Source Assembly
Thermionic Source Assembly Cable