466
TRACE 1600/1610 Hardware Manual
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
5. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
6. Remove the dummy module from the position where the detector module will be
installed.
IMPORTANT The FPD detector should be installed in the back site of the detector
housing to avoid obstacles with another module. For details about the installation of the
FPD detector module, see the section
a. Open the module flap cover.
b. Using a T20 Torxhead screwdriver, unscrew and remove the two captive fixing
screws.
c. Keeping the dummy module flap cover open, lift up the module from its seat in the
injector/detector housing. Place the dummy module on a clean surface.
d. Remove the gas block plug from the gas connections by unscrewing its fixing screw
using a T20 Torxhead screwdriver. See
Figure 462. Detector Gas Block Plug
WARNING Make sure all the four O-rings are placed into their seats on the gas
connection. See
Figure
Do not install the module if the O-rings are missing.
7. Open the front door of the GC.
"Installing the FPD Detector"
Figure
462.
on
page
462.
Gas Block Plug on Detector Seat
O-rings
Gas Connections Without Gas Block Plug
190.
Thermo Scientific