Thermo Scientific TRACE 1600 Hardware Manual page 527

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
6. Remove the dummy module from the position where the detector module will be
installed.
a. Open the module flap cover.
b. Using a T20 Torxhead screwdriver, unscrew and remove the two captive fixing
screws.
c. Keeping the dummy module flap cover open, lift up the module from its seat in the
injector/detector housing. Place the dummy module on a clean surface.
d. Keep the gas block plug of the gas connections installed. See
Figure 493. Detector Gas Block Plug
7. Open the front door of the GC.
8. Plug and fix the detector module into the main frame.
a. Place the module in its seat. Be sure to insert the 25-pin male connector, on the
bottom of the module, into the 25-pin female connector on the detector seat of the
detector housing.
Note Because the module flap cover of the PDD detector cannot be opened, two
holes are provided on the module flap cover for accessing the captive fixing
screws. See
Figure
Figure 494. Holes on the Module Flap Cover
b. Insert a T20 Torxhead screwdriver into the holes provided on the module flap cover
and tighten the captive fixing screws without overtightening.
494.
9
Adding Modules
Adding a PDD Module
Figure
462.
Gas Block Plug on Detector Seat
Holes
TRACE 1600/1610 Hardware Manual
495

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Trace 1610

Table of Contents