Installing The Mounting Bracket - Thermo Scientific TRACE 1600 Hardware Manual

Gas chromatographs
Table of Contents

Advertisement

8
Installing Optional Kits
Installing the NoVent Microfluidics

Installing the Mounting Bracket

402
TRACE 1600/1610 Hardware Manual
Figure 375. Removing the Detector Gas Plug
WARNING Make sure all four O-rings are correctly seated on the gas connection. Do not
install the module if the O-rings are missing.
12. Install the No-Vent module.
a. Open the module flap cover.
b. Keeping the module flap cover open, place the module in its seat. Be sure to insert the
25-pin male connector, on the bottom of the module, into the 25-pin female
connector on the detector seat of the detector housing.
c. Use a T20 screwdriver to tighten the three captive fixing screws without
overtightening.
Note Tighten the center screw first and then secure the side screws.
d. Close the module flap cover.
13. Power on the GC.
14. Switch the module on so that the LED light glows green and start the carrier gas flowing
to purge the line.
Note This module requires a constant carrier gas pressure of 60 psig (410 kPa).
Install the mounting bracket near the front of the GC oven on the right-hand side. This will
keep the mounting bracket out of the way of the column.
Gas Block Plug on Detector Seat
O-rings
Gas Connections Without Gas Block Plug
Thermo Scientific