Figure 54. Adapter Installation on the Bottom of the Detector
a. Finger tighten the retaining nut of the adapter until it starts to grip the bottom of the
detector.
b. Use the 1/2-in. wrench to firmly tighten the retaining nut of the adapter with the
base seal and the silver seal from the bottom of the injector.
9. Connect the column.
a. Before starting, use a 1/4-5/16-in. wrench to remove the column nuts and ferrules
from the base of the injector and detector adapters. See
1
Column Installation Requirements
Silver Seal
FID Graphite
Ferrule Adapter
Figure
55.
TRACE 1600/1610 Hardware Manual
Installation
51