Getting Started

Thermo Scientific
Table 15. Packed Column Adapters and 1/2-in. Wrench
Part
Install the column adapters following the instructions in the next operating procedures.
"To install the packed column adapters"
"To connect a new packed column to the injector and detector adapters"
To install the packed column adapters
1. Put the GC in standby condition.
2. Cool the oven and injector to room temperature.
3. Turn the carrier gas off, and wait for the carrier pressure to go to zero.
Note By pressing the Maintenance button, the GC cool down is automatically
carried out.
4. Remove the analytical column.
a. Open the front door of the GC.
b. Loosen the retaining nut from the injector and detector fitting on the upper interior
wall of the GC oven.
c. Remove the analytical column with its nut and ferrule from the bottom of the
injector and the detector.
Description
Adapter for the connection of the 1/8-in. OD packed column
to the SSL injector
Adapter for the connection of the 1/16-in. OD packed
column to the SSL injector
Adapter for the connection of the 1/8-in. OD packed column
to the FID, NPD, ECD, TCD, or FPD detector
Adapter for the connection of the 1/16-in. OD packed
column to the FID, NPD, ECD, or FPD detector
1/2-in. wrench
on
page 359
8
Installing Optional Kits
Installing the Packed Column Adapters
on
page 362
TRACE 1600/1610 Hardware Manual
359