Installing the Packed Columns Support Kit

Thermo Scientific
This section provides instructions for installing the support kit for packed columns
PN 19070124 into the oven of the TRACE 1600/1610 GC.
Figure 443
shows the support for packed columns.
Figure 443. Support for Packed Columns
Front View
Fixing Slot
Back View
To install the support for packed column kit
1. Put the GC in standby condition.
2. Cool the oven, injectors and detectors to room temperature.
3. Open the front door of the GC.
4. Locate the two slots on the upper section of the oven baffle and the fixing hole on the
ceiling of the oven. See
Removable Rack for
Capillary Column
Figure
444.
8
Installing Optional Kits
Installing the Packed Columns Support Kit
Lock Bridge
Hooks
TRACE 1600/1610 Hardware Manual
447