Thermo Scientific
Carrier line filter
Split line filter
b. Install the dummy filters in their seats interposing the O-ring, then turn them
clockwise.
O-ring
Dummy filter
7. Replace the current liner installed into the Split/Splitless injector with the dedicated
splitless liner.
8. Close the module flap cover.
9. Install the uncoated precolumn.
a. Open the oven door.
b. Disconnect the analytical column from the bottom of the injector.
8
Installing Optional Kits
Installing the Large Volume Splitless Kit
TRACE 1600/1610 Hardware Manual
377