Thermo Scientific TRACE 1600 Hardware Manual page 301

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
8. Remove the top parts of the injector. See
a. Unscrew the septum cap of the injector.
b. Using tweezers, remove the septum from the top of the injector head assembly.
9. Remove the liner. See
a. Using the slotted stubby driver provided, unscrew and remove the liner cap.
b. Use tweezers to remove the liner with the liner seal (O-ring) from the injector.
CAUTION Be careful not to break the glass liner when removing it. Glass splinters
might fall into the lower part of the vaporization chamber. If the glass liner
breaks, follow the instructions for
page
144.
ATTENTION Veillez à ne pas briser le liner en verre pendant son extraction. Des
éclats de verre pourraient tomber dans la partie inférieure de la chambre de
vaporisation. En cas de rupture du liner en verre, consultez
liner PTVBKF
Figure
Figure
234.
"Replacing the PTVBKF Broken Liner"
cassé".
6
Injectors Advanced Maintenance
Cleaning the PTVBKF Injector Head Assembly
234.
"Remplacement d'un
TRACE 1600/1610 Hardware Manual
on
269