Thermo Scientific TRACE 1600 Hardware Manual page 555

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
Figure 531. Bending of the Detector Gas Tubing on the TRACE 1600 Auxiliary Oven
3. Connect the relevant detector gas tubing to the manifold of the GDI electrical interface.
See
Figure
532.
Figure 532. Detector Gas Tubing Connection to the GDI Electrical Interface - External View (1)
a. Guide the detector gas tubing up to reach the GDI electrical interface.
b. Bend and run the detector gas tubing along the back panel of the GC (or TRACE
1600 Auxiliary Oven) until its end reaches the corresponding numbered gas outlet
port on the GDI electrical interface.
c. Connect the detector gas tubing to the corresponding numbered outlet port using the
appropriate Swagelok® 1/16-in. nut and ferrules. Use a 1/4-in. wrench to tighten the
fittings.
d. Repeat
step b
and
the GDI electrical interface.
The result of the operation is shown in
Third-party
Detector
step c
until all the required detector gas tubing are connected to
Figure
533.
9
Adding Modules
Adding a Generic Detector Interface
Gas Tubing
TRACE 1600/1610 Hardware Manual
Gas 3
Gas 2
Gas 1
523