Thermo Scientific TRACE 1600 Hardware Manual page 175

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
11. Reinstall the top parts of the injector.
a. Place the septum into the injector head assembly.
CAUTION It is suggested to replace the septum with a new one.
ATTENTION Il est recommandé de remplacer le septum par un modèle neuf.
b. Screw and tighten the septum cap to finger-tight.
CAUTION Do not overtighten the septum cap because you might damage the septum and
affect instrument performance.
ATTENTION Ne serrez pas excessivement le capuchon du septum, car cela pourrait
endommager le septum et affecter les performances de l'instrument.
12. Close the module flap cover.
13. If present, move the autosampler towards the module to restore the original alignment.
14. Turn the carrier gas on.
15. Set the normal injector, detector, and GC working conditions.
Maintaining a Programmable Temperature Vaporizing Injector with Backflush (PTVBKF)
3
Performing Injectors Routine Maintenance
TRACE 1600/1610 Hardware Manual
143