Thermo Scientific
5. Move the Tee connector support on the lower part of the mounting bracket, then
finger-tighten the nut up to grip the clamp without overtightening. This allow to adjust
the position of the Tee connector during the installation of backflush line, precolumn,
and capillary column.
6. At this point mount the mounting bracket into the oven of the GC. See
mounting bracket"
on
To install the mounting bracket into the GC oven
Install the mounting bracket near the front of the GC oven on the right-hand side. This will
keep the mounting bracket out of the way of the column.
Proceed as follows:
1. Make sure that the nut at the top end of the mounting bracket are removed. See
Figure 359
on
page
383.
2. Line up the bottom of the mounting bracket into two holes on the bottom side of the GC
oven.
3. Hold the mounting bracket directly upright, and twist the top end of it so that it is
securely attached to the GC oven holes.
4. Let out the top of the mounting bracket until it is long enough to be secured into two
holes on the bottom of the GC oven.
5. Replace the nut.
6. Tighten the nut until the mounting bracket is securely attached to the top and bottom of
the GC oven.
Figure 363
shows the result of the operation.
page
390.
8
Installing Optional Kits
Connecting a SSL/PTV Backflush System
"To install the
TRACE 1600/1610 Hardware Manual
385