Thermo Scientific TRACE 1600 Hardware Manual page 120

Gas chromatographs
Table of Contents

Advertisement

3
Performing Injectors Routine Maintenance
Maintaining a Split/Splitless Injector (SSL)
88
TRACE 1600/1610 Hardware Manual
b. Using tweezers, remove the septum from the septum holder.
c. Avoid touching the septum with your fingers. Insert a new septum into the septum
holder using tweezers.
d. Screw and tighten the septum cap to finger-tight.
CAUTION Do not overtighten the septum cap because you might damage the septum and
affect instrument performance.
ATTENTION Ne serrez pas excessivement le capuchon du septum, car cela pourrait
endommager le septum et affecter les performances de l'instrument.
7. Close the module flap cover.
8. If present, move the autosampler towards the module to restore the original alignment.
9. Turn the carrier gas on.
10. Set the normal injector, detector, and GC working conditions.