Thermo Scientific
2. Remove the left side panel.
a. Open the front door of the GC.
b. Use a T20 Torxhead screwdriver to screw the left side panel screw from the interior
front panel. See
Figure
Figure 615. Left Side Panel Fixing Screw
c. Slide the panel towards the back of the instrument up to the stop.
d. Remove the panel pulling it outward being aware that the ground wire is attached to
the panel. See
Figure
e. Unplug the ground wire from the panel.
Figure 616. Left Side Panel Removal
3. Prepare the duct for the transfer line inner tube.
Upgrading a Stand Alone TRACE 1600/1610 to MS Version
615. Save the screw because it will be reused later.
616.
11
Upgrading Equipment
TRACE 1600/1610 Hardware Manual
591