Detectors Advanced Maintenance
Cleaning or Replacing the FPD Filter-side Heat Shields
322
TRACE 1600/1610 Hardware Manual
CAUTION If the filter installed is equipped with the spacer (see the kits PN 19050785 and
PN 19050786), first insert the filter, and then the spacer.
Interferential Filter
Spacer
ATTENTION Si le filtre installé est équipé d'un espaceur (voir les kits réf. 19050785 et
réf. 19050786), insérez d'abord le filtre, puis l'espaceur.
b. Reassembly the photomultiplier assembly and the detector body, then fix them
together tightening the knurled nut. See
Figure 296. Photomultiplier Assembly Remounting
11. Reinstall the FPD detector on the base.
a. Place the detector on its base body, ensuring that the Aluminum ring has been
inserted in the correct position, then tighten the fixing nut. See
Figure
296.
Figure
297.