Thermo Scientific TRACE 1600 Hardware Manual page 61

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
Figure 33. Left/Right Interior Oven Wall Box View
d. Place the left/right panel for MS and attach the screw holding it in place.
3. Introduce the transfer line inner tube into the oven through the duct provided.
4. Attach the transfer line to the GC column using the proper nut and ferrule.
CAUTION - INSTRUMENT DAMAGE: Make sure that the GC column has been
conditioned before connecting it to the transfer line. The material released from the
column (column bleed) during conditioning may contaminate the detector.
ATTENTION – RISQUE D'ENDOMMAGEMENT DE L'INSTRUMENT : veillez à ce que la
colonne GC ait été conditionnée avant de la brancher sur la ligne de transfert. Le matériau
libéré par la colonne (bleeding de la colonne) pendant le conditionnement risque de
contaminer le détecteur.
5. Close the front door of the GC.
6. Connect the transfer line heater if required.
a. Connect the transfer line heater to the Aux Temperature/Cryo Module. See
Figure
34.
Left Oven Wall
Right Oven Wall
TRACE 1600/1610 Hardware Manual
1
Installation
Coupling to a Mass Spectrometer
29

Hide quick links: