Thermo Scientific
a. Completely built your gas supply lines including any traps, tees, and extra tubing to
allow about 40 cm (16 in.) of slack in the line.
b. Purged the gas line after every tube cut to remove any debris or contaminants.
c. Ensured the gas supply is turned off.
2. Connect the gas lines.
a. TRACE 1600/1610 is provided with six gas inlet ports for the connection of carrier
and detector gases. See
Figure 4.
Gases Inlet Connections
Carrier Front
Carrier Back
Hydrogen
Air
Make-up Front
Make-up Back
Note Use the 1/8-in. Swagelok fittings provided on the gas inlet ports (see
to connect the gas lines.
Figure 5.
Fittings for Gas Inlets Connection
Nut
Figure
4.
Ferrules
Gas Inlet Fitting
1
Making the Gas Supply Plumbing Connections
O-Ring
TRACE 1600/1610 Hardware Manual
Installation
Figure
5)
9