8
Installing Optional Kits
Installing the Manual On/Off Valve for Single Gas Line
380
TRACE 1600/1610 Hardware Manual
Use a 7/16-in. wrench for unscrewing the fittings.
Gas supply tube
5. Screw the gas inlet fitting 1/8-in. provided into the gas inlet port interposing the O-ring.
Use a 7/16-in. wrench for tightening the fitting.
6. Using the nut and ferrules provided, connect the valve outlet to the gas inlet fitting
1/8-in. Make sure that the arrow marked on the valve body indicating the direction of the
flow is turned forward the GC gas inlet. Use a 7/16-in. wrench for tightening the fittings.
Gas inlet fitting 1/8-in.
Nut and ferrules
Gas flow direction
Gas inlet fitting
O-ring
Thermo Scientific