Thermo Scientific TRACE 1600 Hardware Manual page 380

Gas chromatographs
Table of Contents

Advertisement

8
Installing Optional Kits
Installing the Merlin Microseal High Pressure Valve Kit
348
TRACE 1600/1610 Hardware Manual
To install the Merlin Microseal Valve kit on the SSL/SSLBKF and HeS-S/SL injector
1. Put the GC in standby condition.
2. Cool the oven and injector to room temperature.
3. Turn the carrier gas off, and wait for the carrier pressure to go to zero.
Note By pressing the Maintenance button, the GC cool down is automatically
carried out.
4. Put the autosampler away if present.
5. Open the module flap cover.
6. Remove the septum.
Figure 327. SSL/SSLBKF and HeS-S/SL Injector: Septum Replacement
Septum Cap
Septum
Septum Holder
a. Unscrew and remove the septum cap.
b. Using tweezers, remove the septum from the septum holder.
7. Install the Microseal valve on the injector.