Thermo Scientific TRACE 1600 Hardware Manual page 449

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
Figure 401. Back Column and Front Column Icons
3. The GC Configuration (Front or Back Column) screen opens. Click the Settings tab to
access the post column settings.
Figure 402. Configuring the Post Column Settings
4. Select Yes from the Post Column? menu. Enter the post column length and ID in the
boxes below.
IMPORTANT For hydrogen, the Post Column Length is 1.2 m. For helium, the Post
Column Length is 0.6 m. The Post Column ID is 0.17 mm for both hydrogen and
helium.
5. Click the home icon when you are finished.
To enter the post-column settings on a TRACE 1600
1. Open Chromeleon Configuration window by clicking the icon on your desktop.
2. In the Front Inlet or Back Inlet tab, select PostColumn for Backflush Type.
8
Installing Optional Kits
Installing the NoVent Microfluidics
TRACE 1600/1610 Hardware Manual
417