6
Injectors Advanced Maintenance
Cleaning the HeS-S/SL Injector Body
258
TRACE 1600/1610 Hardware Manual
10. Remove the body head O-rings. See
a. Using tweezers, remove both the internal and external body head O-rings.
b. Place and keep all the removed components on a clean surface.
11. Remove the injector body. See
a. Using a T20 Torxhead screwdriver, undo the two injector body fixing screws, and
extract the injector body from its housing.
Note Do not remove the carrier, split and purge lines O-rings.
Figure 226. Instant Connect Helium Saver Injector Module: Cleaning the Injector Body (3)
Helium Saver SSL Injector Body
12. Clean the injector body.
a. Ultrasonically clean the injector insert using a warm 1% Liquinox™ solution (or
equivalent).
b. Thoroughly rinse the insert then ultrasonically clean in chromatographic grade
methanol or acetone followed by solvents of lower polarity such as methylene
chloride and hexane. Limit ultra-sonication to 5 min in each solvent.
c. Blow dry the insert using high purity gas (do not use compressed house air as it
contains residual oils from the compressor) then assemble in the reverse order of
removal.
Figure
226.
Figure
226.
Helium Transfer Tube
Purge Line O-ring
Split Line O-ring
Carrier Line O-ring
Body Head Internal O-ring
Body Head External O-ring
Thermo Scientific