Thermo Scientific TRACE 1600 Hardware Manual page 471

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
Figure 429. Filter-side Heat Shield Removal
d. Remove with care traces of graphite due to the breaking of the seal.
e. Reinstall the mirror plug.
Assembling the primary FPD detector in Dual FPD configuration
Perform the Dual FPD configuration on the primary FPD detector using the components
removed from the secondary FPD detector.
9. Remove the mirror plug from the cell of the primary FPD detector.
a. Loosen the Allen screws that fix the mirror plug and remove it. See
Figure 430. Mirror Plug Removal
10. Insert the filter support and the flange assembly into the primary FPD detector.
a. Insert the heat shield into its housing inside the detector body.
b. Insert the spacer into the flange, letting it stand out for about 5 mm.
c. Insert the graphite seal on the spacer, pushing it slowly until it touches the flange.
d. Fix the assembly tightening the three Allen screws. See
8
Installing Optional Kits
Performing the Dual FPD Detector Configuration
Figure
Figure
431.
TRACE 1600/1610 Hardware Manual
430.
439

Hide quick links: