Thermo Scientific TRACE 1600 Hardware Manual page 141

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
a. Use tweezers to remove the body head internal and external O-rings, and replace
them with new O-rings.
8. Reinstall the top parts of the injector.
a. Place the septum holder/liner cap with the septum on the body head of the injector
and fix it by screwing the ring nut.
b. Screw and tighten the injector cap to finger-tight.
CAUTION Do not overtighten the septum cap because you might damage the septum and
affect instrument performance.
ATTENTION Ne serrez pas excessivement le capuchon du septum, car cela pourrait
endommager le septum et affecter les performances de l'instrument.
9. Close the module flap cover.
10. If present, move the autosampler towards the module to restore the original alignment.
11. Turn the carrier gas on.
12. Set the normal injector, detector, and GC working conditions.
3
Performing Injectors Routine Maintenance
Maintaining a Split/Splitless Injector with Backflush (SSLBKF)
TRACE 1600/1610 Hardware Manual
109