Detectors Advanced Maintenance
Measuring the NPD Gas Flows
296
TRACE 1600/1610 Hardware Manual
c. Gently lift up the thermionic source connector guiding the flexible cable from the
thermionic source support, then remove the thermionic source assemble from the
NPD body. Avoid bumping the bead on the sides of the collector. See
Figure 260. NPD Thermionic Source Removal (3)
12. Remove the collecting electrode and the insulator. See
a. Using a 6-mm wrench, unscrew and remove the collecting electrode pin from its seat
on the detector body.
b. Using forceps or tweezers, extract the collecting electrode through the top of the
detector body.
c. Using forceps or tweezers, extract the insulator through the top of the detector body.
Place the insulator on a clean surface.
Figure
Thermionic Source Connector
Thermionic Source Flexible Cable
Thermionic Source Assemble
Thermionic Source Support
NPD Body
Collecting Electrode Pin
Figure
261.
Thermo Scientific
260.