Installing the PTV/PTVBKF Cryo System

Thermo Scientific
• Tube for the coolant into the PTV/PTVBKF injector module.
• Coolant tank tube with connection fittings.
• Aux Temperature/Cryo Module that should be installed into a free external module
housing provided on the back of the GC. See
Figure 565. External Modules Housing
To install the PTV/PTVBKF Cryo System
1. Put the GC in standby condition.
2. Cool the oven, injectors and detectors to room temperature.
Note By pressing the Maintenance button, the GC cool down is automatically
carried out.
3. Close the gas supplies.
4. Power off the GC.
a. Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
5. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
6. Put the autosampler away if present.
7. Install the Aux Temperature/Cryo Module.
Adding the PTV and PTVBKF Cryo System
Figure
565.
External Modules Housing
TRACE 1600/1610 Hardware Manual
10
Adding Systems
549