Thermo Scientific TRACE 1600 Hardware Manual page 96

Gas chromatographs
Table of Contents

Advertisement

1
Installation
iConnect Column Lock
64
TRACE 1600/1610 Hardware Manual
Table 11. Column Insertion Depth for SSL, SSLBKF, and GSV Injectors
Injector
SSL
SSLBKF
GSV
Table 12. Column Insertion Depth for ECD, FID, FPD, NPD, and TCD Detectors
Detector
ECD
FID
FPD
NPD
TCD
Figure 66. SSL and SSLBKF Column Depths
CAUTION Use the type of Graphite/Vespel® ferrules provided in the connector outfits.
Ensure ferrules are the appropriate size for the type of capillary column in use.
ATTENTION Utilisez le type de ferrules Graphite / Vespel™ fourni dans les raccords de
branchement. Veillez à ce que les ferrules soient d'une taille appropriée pour le type de
colonne capillaire utilisé.
Note For typical 0.1 to 0.25 mm inner diameter columns with a 0.36 mm outer
diameter, ferrule packages are required with P/N 290VA191 to seal properly. These
ferrules have a hole size of 0.37 mm and a label with "r2" after Pk 10.
12. Turn the iConnect Column Lock a quarter turn until it stops. See
Column Insertion Depth
5 mm (splitless)
10 mm (split)
5 mm (splitless)
10 mm (split)
23 mm
Column Insertion Depth
23 mm
36 mm
125 mm
32 mm
10 mm
Figure
67.
Thermo Scientific