Installing External Accessories

Installing External Accessories

Installing the Injector and Detector Modules

4
TRACE 1600/1610 Hardware Manual
For additional information, refer to the TRACE 1600/1610 User Guide and to the
documentation included with the accessories you purchased for your GC.
This section provides the instructions for installing your front/back injector and detector
modules into the relevant housing.
The GC is shipped with dummy modules installed into the injector/detector housings.
See
Figure
1.
Figure 1.
Injector and Detector Housings and Dummy Modules
Detector Housing
Note Where a dummy module is installed, the gas connection is blocked by a plug.
The dummy modules must be removed and replaced by the injector and detector modules as
required by the configuration of your TRACE 1600/1610.
CAUTION Each housing must always be occupied by both modules. If the GC is
configured with a single injector/detector module, a dummy module must remain
inserted instead of the missing injector/detector module.
The injector and detector modules are shipped with a plug. Before installing the column,
remove the plug.