Figure 222. SSLBKF Injector Components
Septum Cap
Ring Nut
Septum Holder/Liner Cap with Septum
Liner Seal
Liner
Bottom Fittings
9. Remove the body head O-rings. See
a. Using tweezers, remove both internal and external body head O-rings.
b. Place and keep all the removed components on a clean surface.
10. Remove the injector body. See
a. Using a T20 Torxhead screwdriver, undo the two injector body fixing screws, and
extract the injector body from its housing.
Note Do not remove the carrier, split, and purge lines' O-rings.
Base Seal
Washer
Retaining Nut
Ferrule
Capillary Column Retaining Nut
Figure
223.
Figure
223.
6
Injectors Advanced Maintenance
Cleaning the SSLBKF Injector Body
TRACE 1600/1610 Hardware Manual
253