Thermo Scientific TRACE 1600 Hardware Manual page 365

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
Figure 313. FPD Detector Reinstallation
b. Reconnect the signal, excitation voltage and ignition/heating cables to the detector.
See
Figure
314.
Figure 314. FPD Cables Connection
13. If other external modules are present, plug the power cable to the AC Input connector of
each module, and to the wall outlet.
7
Detectors Advanced Maintenance
Cleaning or Replacing the FPD Flame-side Heat Shields
TRACE 1600/1610 Hardware Manual
333