Figure 521. GC Backplane Board Layout
b. Connect the detector cable to the connector J13 DET.FRONT, or J5 DET.BACK
accordingly.
ATTENTION TRACE 1600/1610 coupled with a TRACE 1600 Auxiliary Oven
If you are installing the GDI mechanical module in the Front or Back site of a TRACE
1600/1610 GC coupled with a TRACE 1600 Auxiliary Oven, the detector cable of the
GDI electrical interface must be connected as follows:
• Mechanical module installed in the Front site: On the backplane board disconnect
the flat cable from the connector marked J13 DET.FRONT. Connect the detector
cable to the connector J13 DET.FRONT.
• Mechanical module installed in the Back site: Do NOT DISCONNECT the flat
cable from connector marked J5 DET.REAR. Connect the detector cable directly to
the additional connector on the flat cable.
4. Reconnect the ground wire to the back cover terminal.
5. Replace the back cover of the GC proceeding in the reverse order in which it was
removed.
The result of the installation is shown in
Adding a Generic Detector Interface
J5 DET.REAR Connector
J13 DET.FRONT Connector
Figure
522.
TRACE 1600/1610 Hardware Manual
9
Adding Modules
515