Adding Modules
Adding an Aux Temperature/Cryo Module
480
TRACE 1600/1610 Hardware Manual
2. Cool the oven, injectors, and detectors to room temperature.
Note By pressing the Maintenance button, the GC cool down is automatically
carried out.
3. Close the gas supplies.
4. Power off the GC.
a. Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
5. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
6. Remove the cover of the external modules housing where installing the module.
See
Figure
477.
Figure 477. Housing Cover Removal
a. Using a T20 Torxhead screwdriver, unscrew and remove the left or right housing
cover screws.
b. Remove the covers from the housing.
7. Install the module into the housing
a. Loosen the two hexagonal screws under the module. See
Figure 478. Module Installation (1)
Left Cover
Screws
Right Cover
Screws
Left/Right Housing Covers
Figure
478.
Hexagonal Screws