GC Main Frame Advanced Maintenance
Removing/Replacing the GC Left Side Panel
212
TRACE 1600/1610 Hardware Manual
a. Open the front door of the GC. Using a T20 Torxhead screwdriver, unscrew the left
side panel screw from the interior front panel. See
Figure 185. Left Side Panel Fixing Screw
b. Slide the panel towards the back of the instrument up to the stop.
c. Remove the panel pulling it outwards being aware that the ground wire is attached to
the panel. See
Figure
Figure 186. Left Panel Removal
d. Unplug the ground wire from the panel.
Note Pay attention to the positioning of the ground wire plug, so it can be
reconnected in the same way it was removed.
186.