Detectors Advanced Maintenance
Cleaning or Replacing the FPD Filter-side Heat Shields
320
TRACE 1600/1610 Hardware Manual
8. Remove the filter-side heat shield.
a. Loosen the three Allen screws that fix the filter support to the flange of the detector
body and remove it. See
Figure 291. Filter-side Heat Shield Removal (1)
b. Turn down slowly the detector body and let the filter-side heat shield come out from
the flange. Pay attention not to lose the O-ring inserted in the filter support.
Figure 292. Filter-side Heat Shield Removal (2)
9. Insert a new filter-side heat shield.
a. Insert a new filter-side heat shield into its housing inside the filter support.
See
Figure
293.
Figure 293. Filter-side Heat Shield Replacement (1)
Figure
291.
Flange
Filter Support
Filter-Side Heat Shield
®
Viton
Thermo Scientific