Installation
Coupling to a Mass Spectrometer
Setting Handshake Parameters
30
TRACE 1600/1610 Hardware Manual
Figure 34. Transfer Line MS Heaters
7. Tune and set the DFS, IRMS, and ICP-MS working conditions of the high resolution
mass spectrometer, referring to the relevant manuals.
When your GC is connected to a mass spectrometer, set the handshaking parameters as shown
in
Table
1.
Table 1. GC Handshaking Parameters When A Mass Spectrometer is Connected
Parameter
Remote Start In
Inhibit Ready In
End of Run Out
Start of Run Out
GC Ready Out
Prep Run Out
To configure handshaking parameters through the touch screen
1. In the main menu select the Configuration icon, the Configuration menu appears.
2. In the Configuration menu, select the Handshake icon to open the relevant submenu.
3. Set the Handshaking parameters as required, then return to main menu.
To configure handshaking parameters through the Chromatography Data System
1. Launch the Data System. In the relevant Configuration page specify the handshaking
parameters.
Heater Connection for
Transfer Line 2
Heater Connection for
Transfer Line 1
Set To:
High to Low
When High
High to Low
High to Low
When Low
When Low