b. Place the injector module in its seat. Be sure to insert the 25-pin male connector, on
the bottom of the module, into the 25-pin female connector on the injector seat of
the injector housing.
c. Use a T20 Torxhead screwdriver to tighten the three captive fixing screws without
overtightening.
IMPORTANT To maintain the correct alignment the screws must be tightened in turn.
Tighten each screw only a small amount before moving to the next screw. Repeat until all
are secure.
9. Remove the top parts of the injector. See
Figure 348. Injector Top Parts Removal
Septum Cap
Ring nut
Septum holder/Liner cap with septum
a. Unscrew the septum cap of the injector.
b. Unscrew the ring nut.
c. Remove the septum holder/liner cap with septum from the injector body head.
10. Replace the liner.
IMPORTANT We suggest replacing the liner currently installed into the injector with
the HS/SPME straight glass empty liner PN 453A1335.
a. Use tweezers to remove the current liner with the liner seal (O-ring) from the
injector.
CAUTION Be careful not to break the glass liner when removing it.
Glass splinters might fall into the lower part of the vaporization chamber.
Installing the HS Adapter Kit on the SSL/SSLBKF Injector
Figure
348.
TRACE 1600/1610 Hardware Manual
8
Installing Optional Kits
367