Installing Optional Kits
Installing the HS Adapter Kit on the SSL/SSLBKF Injector
366
TRACE 1600/1610 Hardware Manual
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
5. Remove the column end from the injector and the detector.
a. Open the front door of the GC.
b. Loosen the retaining nut from the injector fitting on the upper interior wall of the
GC oven.
c. Remove the analytical column with its nut and ferrule from the bottom of the
injector.
6. Remove the SSL/SSLBKF injector module from its seat.
a. Open the module flap cover.
b. Using a T20 Torxhead screwdriver, unscrew the three captive fixing screws.
c. Lift the module from its seat in the injector housing. Place the SSL/SSLBKF module
on a clean surface.
WARNING Make sure the O-ring remains into its seat on the gas connection.
Do not install the module if the O-ring is missing.
7. Replace the module flap cover with the one's provided. See
Figure 347. Module Flap Cover for the HS Adapter
8. Reinstall the injector module into the main frame.
a. Open the module flap cover.
O-ring
Figure
347.
Module flap cover
Thermo Scientific