Thermo Scientific TRACE 1600 Hardware Manual page 281

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
Figure 220. SSL Injector Components
Septum Cap
Ring Nut
Septum Holder/Liner Cap with Septum
Liner Seal
Liner
Base Seal
Washer
Retaining Nut
Ferrule
Capillary Column Retaining Nut
9. Remove the body head O-rings. See
a. Using tweezers, remove both the internal and external body head O-rings.
b. Place and keep all the removed components on a clean surface.
10. Remove the injector body. See
a. Using a T20 Torxhead screwdriver, undo the two injector body fixing screws, and
extract the injector body from its housing.
Figure
221.
Figure
221.
TRACE 1600/1610 Hardware Manual
6
Injectors Advanced Maintenance
Cleaning the SSL Injector Body
Bottom Fittings
Terminal Fitting for Capillary Column
249

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Trace 1610

Table of Contents