Thermo Scientific TRACE 1600 Hardware Manual page 44

Gas chromatographs
Table of Contents

Advertisement

1
Installation
Making the Gas Supply Plumbing Connections
12
TRACE 1600/1610 Hardware Manual
Figure 11. TCD Detector Gas Lines Connection
Figure 12. FPD Detector Gas Lines Connection
f.
Be sure to complete all inlet and detector connections before turning on the gas
supply.
3. Supply the gas lines.
IMPORTANT The maximum nominal inlet pressure for all the inputs is 1050 kPa
(152 psig), as indicated on the label under the gas inlets ports on the back of the GC.
The working inlet pressure range is from 400 kPa (58 psig) to 1050 kPa (152 psig).
Note If auxiliary carrier gases are required, the installation of the Auxiliary Gas module is
required. For the installation details, see the section
page
558.
Front Carrier Gas Inlet
Front Carrier Gas Inlet
Back Carrier Gas Inlet
Back Carrier Gas Inlet
"Adding an Auxiliary Gas System"