Column Installation Requirements
50
TRACE 1600/1610 Hardware Manual
Figure 52. Adapter Installation on SSL Injector Bottom
Retaining Nut
a. Finger tighten the retaining nut of the adapter until it start to grip the bottom of the
injector.
b. Use the 1/2-in. wrench to firmly tighten the retaining nut of the adapter with the
base seal and the silver seal from the bottom of the injector.
7. Remove the bottom parts of the FID, NPD, ECD, TCD, or FPD detector. See
Figure 53. Detector Bottom Parts Removal
Note Save the bottom parts of the detector in a safe place because they will be reused
when you restore the original configuration.
a. Using the 1/2-in. wrench provided, unscrew the retaining nut from the bottom of the
detector.
8. Install the adapter on the bottom the detector. See
Silver Washer
SSL Bottom
SSL Washer
Inlet
Column Adapter
Nut
Detector Bottom
Retaining Nut
Figure
54.
Figure
53.