Figure 528. Detector Gas Tubing Manifolds (1)
Third-party
Detector
1. Take one, two, or three segments of the 1/16-in.stainless steel tubing (provided)
according to the detector gases required and long enough to properly connect both the
detector gas tubing manifolds.
2. Connect the detector gas tubing to the manifold of the GDI mechanical module.
See
Figure
529.
Figure 529. Detector Gas Tubing Connections
Third-party
Detector
a. Connect the detector gas tubing to the numbered inlet ports using the appropriate
Swagelok® 1/16-in. nut and ferrules. Use a 1/4-in. wrench to tighten the fittings.
9
Adding Modules
Adding a Generic Detector Interface
Detector Gas Tubing
Manifolds
Detector Gases
Tubing
Detector Gases
Tubing Manifold
TRACE 1600/1610 Hardware Manual
521